AD5522
Data Sheet
Rev. E | Page 26 of 64
06197-
026
CH1 10.0mV
B
W CH2 50.0mV
B
W
M100s
1
4
2
3
T
200.000s
TA = 25°C
FOHx VICTIM
MEASOUTx VICTIM
MEASOUTx
ATTACK
TRIGGER
CH2 Pk-Pk 14.38mV
CH4 5.00V
CH3 5.00V
Figure 34. Shorted DUT AC Crosstalk, Victim PMU in FVMI Mode
(±200 μA Range)
1.80
1.75
1.70
1.65
1.60
1.55
1.50
1.45
1.40
25
45
35
65
55
75
85
06197-
127
MEA
SO
U
T
x
VO
L
T
A
G
E
(V)
FORCED TEMPERATURE (°C)
NOMINAL SUPPLIES
±15.25V
5 DIFFERENT DEVICES
Figure 35. Temperature Sensor Voltage on MEASOUTx
vs. Forced Temperature
06197-
128
CH1 100mV
B
W
M2.00s
3
4
1
T
6.0000s
CH3 5.00V BW
CH4 5.00V BW
CH4
2.10V
FOH0
BUSY
SYNC
Figure 36. Range Change, PMU0, ±5 μA to ±2 mA, CLOAD = 1 nF,
RLOAD = 620 kΩ, FV = 3 V
06197-
129
CH1 50.0mV
B
W
M800ns
4
1
3
T
2.40000s
CH3 5.00V BW
CH4 5.00V BW
CH4
2.10V
FOH0
BUSY
SYNC
Figure 37. Range Change, PMU0, ±2 mA to ±5 μA, CLOAD = 1 nF,
RLOAD = 620 kΩ, FV = 3 V
06197-
130
CH1 20.0mV
B
W
M20.0s
4
1
3
T
60.0000s
CH3 5.00V BW
CH4 5.00V BW
CH4
2.10V
FOH0
BUSY
SYNC
Figure 38. Range Change, PMU0, ±5 μA to ±2 mA, CLOAD = 100 nF,
RLOAD = 620 kΩ, FV = 3 V
06197-
131
CH1 20.0mV
B
W
M20.0s
4
1
3
T
60.0000s
CH3 5.00V BW
CH4 5.00V BW
CH4
2.10V
FOH0
BUSY
SYNC
Figure 39. Range Change, PMU0, ±2 mA to ±5 μA, CLOAD = 100 nF,